• LOGIN
    Login with username and password
Repository logo

BORIS Portal

Bern Open Repository and Information System

  • Publications
  • Theses
  • Research Data
  • Projects
  • Organizations
  • Researchers
  • More
  • Collections
  • Statistics
  • LOGIN
    Login with username and password
Repository logo
Unibern.ch
  1. Home
  2. Publications
  3. Technical assessment of the NDI Polaris Vega optical tracking system.

Technical assessment of the NDI Polaris Vega optical tracking system.

Details
Files
DOI
10.48350/156515
Publisher DOI
10.1186/s13014-021-01804-7
PubMed ID
33980248
Abstract
The Polaris product line from Northern Digital Inc. is well known for accurate optical tracking measurements in research and medical environments. The Spectra position sensor, to date often found in image guided radiotherapy suites, has however reached its end-of-life, being replaced by the new Vega model. The performance in static and dynamic measurements of this new device has been assessed in controlled laboratory conditions, against the strict requirements for system integration in radiation therapy. The system accuracy has improved with respect to the Spectra in both static (0.045 mm RMSE) and dynamic (0.09 mm IQR, < 20 cm/s) tracking and brings marginal improvement in the measurement latency (14.2 ± 1.8 ms). The system performance was further confirmed under clinical settings with the report of early results from periodic QA tests within specifications. Based on our tests, the Polaris Vega meets the quality standards of radiotherapy applications and can be safely used for monitoring respiratory breathing motion or verifying patient positioning.
Date Issued
2021-05-12
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Subjects
Breathing motion IGRT Optical tracking Patient positioning
Language(s)
en
Author(s)
Fattori, Giovanni
Lomax, Antony John
Weber, Damien Charles  
Universitätsklinik für Radio-Onkologie  
Safai, Sairos
Additional Credits
Universitätsklinik für Radio-Onkologie  
Journal
Radiation oncology
Publisher
BioMed Central
ISSN
1748-717X
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 24f0a9 [ 4.09. 8:55]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify