Implant-supported fixed complete denture frameworks in additively and subtractively manufactured high-performance polymers: Fabrication and fit accuracy.
Publisher DOI
PubMed ID
40409702
Abstract
Objective
To compare the fabrication and fit accuracy of additively (AM) and subtractively (SM) manufactured implant-supported fixed complete denture (ISFCD) frameworks in high-performance polymers with those in titanium.Methods
An edentulous maxillary model with four implants (right first molar, right canine, left canine, and left right first molar) was digitized to design an ISFCD framework, which was used to fabricate a total of 50 frameworks in polyetheretherketone (AM-PEEK and SM-PEEK), polyetherketoneketone (AM-PEKK and SM-PEKK), and titanium (SM-Ti) (n = 10). A laboratory scanner was used to digitize each framework and when the frameworks were seated on the model at the left first molar site with the 1-screw test. The scan files of the frameworks were superimposed over the design file to evaluate the surface and interimplant deviations, while the fit at right first molar, right canine, and left canine sites was assessed with the average gap values. Data were statistically analyzed (α = 0.05).Results
SM-PEEK and SM-Ti frameworks mostly had higher and AM frameworks mostly had lower surface accuracy (P ≤ 0.036). SM-Ti and SM-PEEK frameworks had the lowest molar-to-molar interimplant distance deviations, and AM frameworks had the lowest precision of interimplant distances (P ≤ 0.014) AM-PEKK frameworks mostly had lower average gap accuracy regardless of the abutment site, followed by AM-PEEK (P < 0.001).Conclusions
SM-PEEK and SM-Ti frameworks mostly had higher and AM frameworks mostly had lower accuracy. AM frameworks had larger marginal gaps. Tested frameworks tended to distort towards the midline.
To compare the fabrication and fit accuracy of additively (AM) and subtractively (SM) manufactured implant-supported fixed complete denture (ISFCD) frameworks in high-performance polymers with those in titanium.Methods
An edentulous maxillary model with four implants (right first molar, right canine, left canine, and left right first molar) was digitized to design an ISFCD framework, which was used to fabricate a total of 50 frameworks in polyetheretherketone (AM-PEEK and SM-PEEK), polyetherketoneketone (AM-PEKK and SM-PEKK), and titanium (SM-Ti) (n = 10). A laboratory scanner was used to digitize each framework and when the frameworks were seated on the model at the left first molar site with the 1-screw test. The scan files of the frameworks were superimposed over the design file to evaluate the surface and interimplant deviations, while the fit at right first molar, right canine, and left canine sites was assessed with the average gap values. Data were statistically analyzed (α = 0.05).Results
SM-PEEK and SM-Ti frameworks mostly had higher and AM frameworks mostly had lower surface accuracy (P ≤ 0.036). SM-Ti and SM-PEEK frameworks had the lowest molar-to-molar interimplant distance deviations, and AM frameworks had the lowest precision of interimplant distances (P ≤ 0.014) AM-PEKK frameworks mostly had lower average gap accuracy regardless of the abutment site, followed by AM-PEEK (P < 0.001).Conclusions
SM-PEEK and SM-Ti frameworks mostly had higher and AM frameworks mostly had lower accuracy. AM frameworks had larger marginal gaps. Tested frameworks tended to distort towards the midline.
Date Issued
2025-08
Publication Type
Article
Subject(s)
Subjects
Additive manufacturing
•
Fit
•
Implant-supported complete fixed denture
•
Polyetheretherketone
•
Polyetherketoneketone
Language(s)
en
Author(s)
Küçükekenci, Ahmet Serkan | |
Güven, Mehmet Esad | |
Dede, Doğu Ömür |
Journal
Journal of Dentistry
Publisher
Elsevier
ISSN
1879-176X
0300-5712
Access(Rights)
open.access