Using an orbit shield during volume perfusion CT: is it useful protection or an obstacle?
Hakim, A., Vulcu, S., Dobrocky, T., Z’Graggen, W. J., & Wagner, F. (2018). Using an orbit shield during volume perfusion CT: is it useful protection or an obstacle? Clinical Radiology, 73(9), 834.e1-834.e8. https://doi.org/10.1016/j.crad.2018.05.003
Publisher DOI
PubMed ID
29859633
Abstract
AIM
To evaluate the image quality and artefacts resulting from in-plane orbit shielding during computed tomography (CT) perfusion and to assess the possibility of artefact reduction using the iterative metal artefact reduction (iMAR) algorithm.
MATERIALS AND METHODS
Twenty-eight perfusion CT examinations obtained from 20 patients using orbit shields were included in this study. Source images and perfusion maps were analysed retrospectively to detect the type and extent of artefacts. Original images were compared with images processed using the iMAR algorithm. The extent of artefacts was categorised into three groups: orbital/frontal, middle fossa, and posterior fossa. Types of artefacts in source images were categorised as noise, streak, beam hardening, or a combination of those, and types of artefacts in perfusion maps were categorised as: noise, image distortion, areas with missing colour codes, or incorrect colour-coding.
RESULTS
All source images evaluated showed artefacts related to orbit shielding and 85.7% reached the posterior fossa. In 92.8% of scans, perfusion maps showed artefacts, mostly as incorrect colour codes. Of the perfusion maps with artefacts, 92.3% reached the posterior fossa. After application of the iMAR algorithm, an increase in the extent of artefacts was observed in 93% of perfusion maps, and no improvement of image quality was noted.
CONCLUSION
Orbit shields cause significant artefacts if used for in-plane shielding during whole-brain CT perfusion, and render areas at the level of the shield undiagnosable. Usage of an orbit shield during whole-brain CT perfusion is not recommended, so alternative methods for reducing the radiation dose are advisable.
To evaluate the image quality and artefacts resulting from in-plane orbit shielding during computed tomography (CT) perfusion and to assess the possibility of artefact reduction using the iterative metal artefact reduction (iMAR) algorithm.
MATERIALS AND METHODS
Twenty-eight perfusion CT examinations obtained from 20 patients using orbit shields were included in this study. Source images and perfusion maps were analysed retrospectively to detect the type and extent of artefacts. Original images were compared with images processed using the iMAR algorithm. The extent of artefacts was categorised into three groups: orbital/frontal, middle fossa, and posterior fossa. Types of artefacts in source images were categorised as noise, streak, beam hardening, or a combination of those, and types of artefacts in perfusion maps were categorised as: noise, image distortion, areas with missing colour codes, or incorrect colour-coding.
RESULTS
All source images evaluated showed artefacts related to orbit shielding and 85.7% reached the posterior fossa. In 92.8% of scans, perfusion maps showed artefacts, mostly as incorrect colour codes. Of the perfusion maps with artefacts, 92.3% reached the posterior fossa. After application of the iMAR algorithm, an increase in the extent of artefacts was observed in 93% of perfusion maps, and no improvement of image quality was noted.
CONCLUSION
Orbit shields cause significant artefacts if used for in-plane shielding during whole-brain CT perfusion, and render areas at the level of the shield undiagnosable. Usage of an orbit shield during whole-brain CT perfusion is not recommended, so alternative methods for reducing the radiation dose are advisable.